EvoStim P therapeutic unit: one stimulation channel, one pressure biofeedback channel. Composed of batteries, instruction manual, connection cable and RU-V2STFW probe. Ideal for perineal reeducation
Product in stock. Immediate shipping
* The shipping costs are calculated in the shopping basket
Reference: evoStim-P
Description
EvoStim P therapy unit: electrostimulation, biofeedback (pressure), triggered stimulation (pressure-triggered)
The EvoStim P electrostimulation device is an exclusive electromedical device and offers professional benefits, with the simplicity of therapeutic units. It allows passive treatments (with electrostimulation), active recovery (with biofeedback) and muscular-cognitive re-education (with provoked stimulation). The P model is used for grasping biofeedback and pressure-induced stimulation . It is the most indicated for perineal reeducation .
Strengths:
- Large LCD with touch screen
- 6 quick selection buttons
- Editable preset programs
- Circular visual feedback
- Visual feedback from bars
- Backlit display
- One stimulation channel
- A biofeedback channel
- Stimulation provoked
- The evoStim® P package includes a balloon-shaped vaginal probe for biofeedback and electrodes for stimulation.
Technical characteristics:
- Number of channels: 1 stimulation + 1 biofeedback
- Power supply: 4.5V with 3 1.5V Alkaline AAA batteries (LR03). Battery life: 20 hours.
- Output current: 60mApp at 1000 Ohm load adjustable in 1mA steps
- Frequency: 1 to 150 Hz (1-2-3-4-5-10-12-14-16-18-20-25-30-35-40-45-50-60-70-80-90 -100-110-120-130-140-150) (± 5%)
- Pulse width: adjustable from 50 to 400 μs in steps of 10 (± 5%)
- Type of impulse: selectable between BIPHASIC SYMMETRICAL and BIPHASIC SYMMETRIC ALTERNATES
- Rise time: automatic depending on the frequency during the ACTION
- Holding time: from 1 to 60 sec. in steps of 1 s. (± 1 s.)
- Lowering time: automatic depending on the frequency during the ACTION
- Rest time: from 0 to 60 sec. in steps of 1 s. (± 1 s.)
- Standby frequency: automatic depending on the frequency during the ACTION
- Session time: selectable between 5 - 90 min '. (± 1min '.) Or continuous stimulation (C).
- Connections to applied parts: protected 2 mm male connector
- Dimensions and weight: 140x70x25 mm - 150 g. batteries included
- Environmental limits: 15-35 ° C - 20-80% UR